Table 5 Comparison of workflow efficiencies and throughput on the cobas 4800 and cobas 6800 systems (including p 480 instruments)

From: HPV testing for cervical cancer screening: technical improvement of laboratory logistics and good clinical performance of the cobas 6800 in comparison to the 4800 system

 

p 480 v1 + 4800

p 480 v2 + 6800

Preanalytical step:

 - samples’ aliquoting

performed by × 480

performed by p 480

Analytical steps:

 - instrument loading (reagents)

single-use vials with pouring (10 min)

re-usable reagent cartridges, no preparation needed

 - reagent on-board stability

discarded after run

90 days

 - user interaction

vials loading/unloading at each run

cartridge loading - unloading when empty/expired

 - instrument loading (samples)

94 uncapped primary vials

up to 280 secondary tubes

 - throughput (per 8 h)

192

384

 - time to first 96 results

4.9 h

< 3.5 h

 - time to each additional 96 results

160 min

90 min

 - hands-on-time (8 h)

60 min

30 min

Instrumentation maintenance

daily

monthly

Amount of waste produced

~ 2.6 l of liquids

~ 1.7 l of liquids / ~ 40% less of sol